ZOLL Propaq MD Rev A User Manual

Page 68

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

64

9650-0807-01 Rev. A

6

Disconnect the battery communication
cable (yellow, orange, brown, red)
connector from the monitor board.

7

Disconnect the monitor/defibrillator flex
cable from the monitor board.

Note: Be careful to pull the connector
straight out. Rocking the connector or
disconnecting it at an angle may result in
bent pins.

8

Disconnect the three power cable
connectors (battery (black), AUX (white),
and defibrillator (red)).

9

Press the tabs on either side of the two
printer cables to disconnect them.

10

Use a wooden stick to release the tabs on
either side of brown flex connector between
the dock connector and the monitor board.

Note: Release one side at a time.

battery
cable

flex cable

p

cables

three power
cables

three power
cables

p

three power
cables

three power
cables

printer cables

NIBP
hoses