C

HAPTER

2

T

ROUBLESHOOTING

50

9650-0807-01 Rev. A

CO2 GAS INPUT
OCCLUSION

CO

2

inlet blocked.

Replace FilterLine

®

. If

problem persists, check for
CO

2

inlet occlusion. If

problem still persists,
replace the CO

2

module.

CO2 MAINTENANCE
REMINDER

Periodic maintenance
reminder.

Replace CO

2

module soon.

CO2 MAINTENANCE
REQUIRED

Periodic Maintenance due.

Replace CO

2

module.

CO2 NO FILTERLINE

No filterline is detected.

If FilterLine is connected
and problem persists,
replace FilterLine. If
problem still persists,
replace the CO

2

module.

DEFIB CAP LEAKAGE

Capacitor bank cannot hold
a charge.

Replace the capacitor bank
module.

DEFIB CHARGE ERROR

Over voltage was detected.

If problem persists, replace
the defibrillator board
module.

DEFIB CHARGE TIMEOUT
FAIL

The battery did not produce
enough current to charge
capacitor.

Charge the battery. If
voltages are okay (from
battery and aux), replace
the defibrillator board
module.

DEFIB DEVICE FAILURE

One of the following issues
occurred:

Device could not charge
up the self-test energy.

Discharge circuit failed to
discharge voltage within
specified time duration.

Device failed to deliver
correct self-test energy.

Check communication
cables. If problem persists,
replace defibrillator board
module.

DEFIB DISCHARGE TEMP

Device is in a hot
environment or the
discharge circuit is always
on.

Move device to cooler area.
If problem persists, replace
the defib board module.

DEFIB FATAL CHARGE
ERROR

Charging error.

Replace the defibrillator
board module.

Error Message

Explanation

Technical Action