Safety Precautions
WARNING!
SHOCK HAZARD!
Caution
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS.
•
Only properly trained technicians should service the unit.
•
The unit can contain deadly voltages even if the unit is turned off.
•
Make sure to discharge the unit before working with it.
•
Make sure you take the necessary precautions when working with static sensitive units. For
example, you must wear a conductive wrist strap (which touches your skin) connected to a
grounding mat and to the earth ground. You must remove the wrist strap when you discharge
high voltage or when you are working on energized equipment.
Modules
The Propaq MD unit contains the following nine (9) modules or assemblies:
•
Printer
•
Front Enclosure
•
Display
•
NIBP
•
EtCO
2
(Oridion
®
)
•
Monitor Board
•
Capacitor Bank
•
Defibrillator Board
•
SpO
2
These modules can be purchased as replacement parts from the ZOLL Technical Support