Nibp test – ZOLL Propaq MD Rev A User Manual

Page 46

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

42

9650-0807-01 Rev. A

NIBP Test

The primary purpose of an NIBP simulator is to reproduce a pressure profile similar to a live
patient to be used for testing repeatability and functionality of the system. There are many
different NIBP simulators, and manufacturers use different methods to develop their algorithm,
which can cause varying readings from different simulators. In order to test for repeatability,
you should first establish the offset of your simulator. The offset is a value that is produced on
the NIBP monitor that is shifted from the simulator’s setting. This value must be established
based on the a statistical sample of monitors and readings.

Note:

Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.

Equipment

Dual lumen NIBP hose, NIBP simulator

Do This...

Observe This...

Pass/Fail

1.

On the Propaq MD unit, ensure the patient
mode is set to Adult.

2.

Set the simulator to the 80/50 simulator
preset.

3.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o

4.

Set the simulator to the 120/80 simulator
preset.

5.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o

6.

Set the simulator to the 200/150 simulator
preset.

7.

Take a manual reading.

Confirm that the NIBP measurement
reported is within +/- 10 mmHg of
the expected simulator performance.

o

o

8.

Disconnect hose.

Advertising