Error Message

Explanation

Technical Action

ALARM SILENCE SHORT

Alarm Silence/Reset key is
stuck.

Replace the front panel.

BLUETOOTH PHONE
UNAVAILABLE

The Propaq MD unit was
unable to establish a Wide
Area Network connection.
None of the Bluetooth
phones paired with the
device were detected.

To correct the problem,
verify the following:

•

The Bluetooth device is
paired with the Propaq
MD unit.

•

The phone is on and
near enough to the
device to permit
communication.

•

The phone is configured
to allow data connection
requests (if the phone
has such a configuration
setting).

CELLULAR CONNECT
FAILURE

The Propaq MD unit was
unable to establish a Wide
Area Network connection. A
cellular modem or Bluetooth
cellular phone connection
was attempted, but failed.

To correct the problem,
verify the following:

•

The cellular provider
information, particularly
the call number, is
correctly configured.

•

The cellular device has
an active plan.

•

There is adequate
cellular network signal
coverage in the area.

•

(For phones) The phone
is configured to allow
data connection
requests (if the phone
has such a configuration
setting).

•

(For modems) The
cellular modem is a
supported device and is
securely connected to
the Propaq MD unit.

If all the configurations are
correct, retry the operation.

Advertising