Leakage current test – ZOLL Propaq MD Rev A User Manual

Page 25

Advertising
background image

Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

21

Leakage Current Test

Equipment

See the manufacturer’s instructions or supplied specifications for the leakage tester you use.

Setup

See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.

Maximum Leakage Acceptance Limits

Patient Leakage

Mains on Applied Part

Normal

Single Fault

Single Fault

ECG

10

50

100

MFC

10

50

100

Normal

Single Fault

Earth Leakage

2500

5000

Advertising