Maintenance Tests
9650-0807-01 Rev. A
Propaq MD Service Manual
43
Temperature Test
Equipment
IBP Temperature Simulator (or equivalent)
Do this...
Observe This...
Pass/Fail
1.
Connect the temperature simulator to the
Propaq MD T1 channel.
2.
On the IBP Temp Simulator, Select 98.6
Degree F (3
7 Degree C) or a setting as close
as possible as your simulator allows.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
o
o
3.
Repeat steps 1 and 2 of this procedure for the
T2 channel.
4.
Disconnect cable.