Synchronized cardioversion test – ZOLL Propaq MD Rev A User Manual
Page 31
Maintenance Tests
9650-0807-01 Rev. A
Propaq MD Service Manual
27
Synchronized Cardioversion Test
Equipment
Fluke Impulse 7000DP Analyzer (or equivalent)
Test Setup
Connect the cable to the defibrillator analyzer.
Select Sync mode on analyzer.
Input 1mV ECG signal at 60 BPM.
Do this...
Observe this...
Pass/Fail
1.
Verify that the Propaq MD unit is
turned on and the
PADS
lead is
selected.
2.
Press the SYNC button on the front of
the defibrillator. Enter the
synchronized cardioversion timing test
mode on the defibrillator analyzer.
Sync appears on display.
Sync markers display on the monitor.
The sync marker appears as an S over
the ECG R-wave peaks on strip chart
and display.
o
o
3.
Select 200J.
4.
Press the CHARGE button. When the
Shock button illuminates, press and
hold it.
Observe that the R-wave to shock delay
(sync delay) is less than 60 milliseconds
on the analyzer display.
Defibrillator discharges.
o
o