Propaq MD Error Messages

9650-0807-01 Rev. A

Propaq MD Service Manual

51

DEFIB FIRE FAILURE

High impedance or low
impedance (short).

Replace pads. If problem
persists, check internal
connections. If problem still
persists, replace the
defibrillator board module.

DEFIB LOW BATTERY

Battery is too low to charge
defibrillator.

Charge the battery. Check
battery contacts and clean
if necessary.

DEFIB NO ENERGY
DELIVERED

High impedance. No defib
energy was delivered.

Replace pads. If problem
persists, check internal
connections. If problem still
persists, replace the
defibrillator board module.

DEFIB PACER DEVICE
FAILURE

One of the following issues
occurred:

Error was detected on the
defib/pacer board.

Interface self-test
communication failed.

Pacer interface self-test
failed.

Communication with
pacer hardware failed
after startup.

Check communication
cables. If problem persists,
replace the defibrillator
board module.

DEFIB PHASE FAILURE

Phase I or phase II
waveform dropout.

Replace the defibrillator
board module.

DEFIB USING DEFAULT
CALIBRATION

Calibration data lost.

Replace the defibrillator
board module.

ECG DEVICE FAILURE

ECG software detected a
problem with the ECG data
that is most likely a hardware
issue.

Replace the main board.

FP DEFIB CHARGE
SHORT

Keys are stuck.

Replace the front panel.

FP DEFIB ENERGY DOWN
SHORT

Defib energy Select Down
(Decrease) key is stuck.

Replace the front panel.

FP DEFIB ENERGY UP
SHORT

Defib energy Select Up
(Increase) key is stuck.

Replace the front panel.

FP DEFIB FIRE SHORT

A key or a front panel button
is stuck.

Replace the front panel.

Error Message

Explanation

Technical Action