Removing the Monitor Board and SpO2 Modules
9650-0807-01 Rev. A
Propaq MD Service Manual
75
4
Remove three (3) screws on the left side of
the unit in the illustrated areas.
5
Turn the unit over.
6
Disconnect the SpO
2
ground cable (as
illustrated).
7
Disconnect ECG flex cable connector from
the left side (as illustrated).
8
Disconnect the SpO
2
connector (as
illustrated).
9
Remove the left side of the unit.
shield
Advertising