Heart rate alarm test – ZOLL Propaq MD Rev A User Manual

Page 21

Advertising
background image

Maintenance Tests

9650-0807-01 Rev. A

Propaq MD Service Manual

17

Heart Rate Alarm Test

Equipment

Fluke Impulse 7000DP Analyzer

Do this...

Observe this...

Pass/Fail

1.

Press the Leads quick access key

(

)

on the Propaq MD unit and

select lead II.

II displays next to the Leads quick
access key.

o

o

2.

On the Propaq MD unit:

•

Press the More quick

access key (

) key to

access the second set of
quick access keys.

•

Press the Alarm quick

access key (

).

•

Press the Limits softkey.

The Alarm Settings menu displays.

o

o

3.

On the Propaq MD unit, select HR/
PR Alarm in the Alarm Settings
menu and use the navigation keys
to do the following:

•

Set the Upper Limit Enable

to ON.

•

Set the Lower Limit

Enable to ON.

•

Set the Lower limit to 35.

•

Set the Upper limit to 150.

Use the navigation keys to exit the
Alarm Settings menu.

4.

On the Propaq MD unit, press the
Alarm Cancel quick access key

(

).

The red alarm suspension timer
displays at the top of the screen next
to the message area.

5.

Remove a lead wire from the

ECG

simulator

.

T

he ECG LEAD FAULT message

displays.

o

o

6.

On the Propaq MD unit, press

.

The Alarm Suspend key flashes

.

7.

Reattach ECG Lead wire to

ECG

simulator

.

The ECG LEAD FAULT message no
longer displays.

o

o

8.

Set simulator to 30 BPM or lower.

Heart Rate Value is highlighted and
the alarm tone sounds. The HR/PR
LOW ALARM
message displays.

9.

On the Propaq MD unit, press

.

The Alarm Suspend key flashes.

The

red alarm suspension timer displays
at the top of the screen next to the
message area.

o

o

10.

Set the simulator to 120 BPM.

11.

On the Propaq MD unit, press

.

Alarm is enabled. The alarm suspend
key flashes. The alarm suspension
timer disappears.

o

o

I, II,

III...