Control using the event/action function

Can start/stop computations or clear/reset computed results by an event occurring. This

function distinguishes clear and reset as shown below.

-

Clear

When issued during the computation, the measured data is reset before doing the first

computation.

Computation clear

Computed data

Computation start

Measurement interval

-

Reset
When issued during the computation, the measured data is reset after doing the first

computation.

Measurement interval

Computation start

Computation
reset

Computed data

Alarm setting for the computation channel

Similar to the standard channels, up to 4 alarm values (levels) can be designated for each

computation channel (upper limit alarm / lower limit alarm).

App.1 Computing Equation