Rp start/stop the report priting, Rs reconstruct the system, Rc initialize the set values – Yokogawa DR240 User Manual
Page 74: Ds transfers the setting mode

6-20
IM DR231-11E
0: Start the report
1: Stop the report
Example
Start the report.
DR0
Comments
• Effective only on instruments with the optional
report function.
• Measurement range, date, time cannot be changed
during a report. Copy using the range copy ON
also cannot be executed during a report.
• All report data up to that point are cleared when a
report is started.
• This setting is effective when one of hourly, daily
and monthly reports is ON.
RP
Start/stop the report priting
Mode
Operation mode
Setting
RPp1<terminator>
p1
Start/stop the report printing
0: Start the report printing
1: Stop the report printing
Example
Start the report printing.
RP0
Comments
• Effective only on instruments with the optional
report function.
• Measurement range, date, time cannot be changed
during a report. Copy using the range copy ON
also cannot be executed during a report.
• All report data up to that point are cleared when a
report is started.
• This setting is effective when one of hourly, daily
and monthly reports is ON.
RS
Reconstruct the system.
Mode
Operation mode
Setting
RSp1<terminator>
p1
0 The system is reconstruct.
Example
Reconstruct the system.
RS0
Comments
• Reconstruct the system. Execute this command if
a subunit or a module is newly added or a module
in a slot is replaced.
• Since it takes an indefinite time to execute the RS
command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after
transmitting the RS command. The format of the
ACK status is as shown below.
E0Cr+Lf The received command was normally
processed.
E1Cr+Lf There is an error in the received
command.
• After executing the command, the clock is
initialized to 96/01/01 00:00:00.
RC
Initialize the set values.
Mode
Operation mode
Setting
RCp1<terminator>
p1
0 The set values are initialized.
Example
Initialize the operation mode parameters (measuring
range, unit, alarm, date & time, and moving
average).
RC0
Comments
• Since it takes an indefinite time to execute the RC
command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after
transmitting the RC command. The format of the
ACK status is as shown below.
E0Cr+Lf
The received command was normally
processed.
E1Cr+Lf
There is an error in the received
command.
• After executing the command, the clock is
initialized to 96/01/01 00:00:00.
DS
Transfers the setting mode.
Mode
All modes
Setting
DSp1<terminator>
p1
0 Transfers to operation mode.
1 Transfers to setup mode.
2 Transfers to A/D calibration mode.
Example
Transfer the DR230/DR240 mode to the setup mode.
DS1
Comments
• Since it takes an indefinite time to execute the DS
command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after
transmitting the DS command. The format of the
ACK status is as shown below.
E0Cr+Lf The received command was normally
processed.
E1Cr+Lf There is an error in the received
command.
6.6 Control and Execution Command