Rp start/stop the report priting, Rs reconstruct the system, Rc initialize the set values – Yokogawa DR240 User Manual

Page 74: Ds transfers the setting mode

Advertising
background image

6-20

IM DR231-11E

0: Start the report
1: Stop the report

Example

Start the report.

DR0

Comments

• Effective only on instruments with the optional

report function.

• Measurement range, date, time cannot be changed

during a report. Copy using the range copy ON
also cannot be executed during a report.

• All report data up to that point are cleared when a

report is started.

• This setting is effective when one of hourly, daily

and monthly reports is ON.

RP

Start/stop the report priting

Mode

Operation mode

Setting

RPp1<terminator>

p1

Start/stop the report printing
0: Start the report printing
1: Stop the report printing

Example

Start the report printing.

RP0

Comments

• Effective only on instruments with the optional

report function.

• Measurement range, date, time cannot be changed

during a report. Copy using the range copy ON
also cannot be executed during a report.

• All report data up to that point are cleared when a

report is started.

• This setting is effective when one of hourly, daily

and monthly reports is ON.

RS

Reconstruct the system.

Mode

Operation mode

Setting

RSp1<terminator>

p1

0 The system is reconstruct.

Example

Reconstruct the system.

RS0

Comments

• Reconstruct the system. Execute this command if

a subunit or a module is newly added or a module
in a slot is replaced.

• Since it takes an indefinite time to execute the RS

command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after
transmitting the RS command. The format of the
ACK status is as shown below.
E0Cr+Lf The received command was normally

processed.

E1Cr+Lf There is an error in the received

command.

• After executing the command, the clock is

initialized to 96/01/01 00:00:00.

RC

Initialize the set values.

Mode

Operation mode

Setting

RCp1<terminator>

p1

0 The set values are initialized.

Example

Initialize the operation mode parameters (measuring
range, unit, alarm, date & time, and moving
average).

RC0

Comments

• Since it takes an indefinite time to execute the RC

command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after

transmitting the RC command. The format of the
ACK status is as shown below.

E0Cr+Lf

The received command was normally
processed.

E1Cr+Lf

There is an error in the received
command.

• After executing the command, the clock is

initialized to 96/01/01 00:00:00.

DS

Transfers the setting mode.

Mode

All modes

Setting

DSp1<terminator>

p1

0 Transfers to operation mode.
1 Transfers to setup mode.
2 Transfers to A/D calibration mode.

Example

Transfer the DR230/DR240 mode to the setup mode.

DS1

Comments

• Since it takes an indefinite time to execute the DS

command, return an ACK after processing. On
the controller side, execute the following
processing after receiving an ACK after
transmitting the DS command. The format of the
ACK status is as shown below.
E0Cr+Lf The received command was normally

processed.

E1Cr+Lf There is an error in the received

command.

6.6 Control and Execution Command