5-2

IM DR231-11E

5.1 Command Format

Terminator

Any one of the following forms a terminator.

• CR + LF
• LF
• EOI = True

(If EOI is to be used for a terminator, add EOI = True to the last parameter character.)

Sub-delimita

Several commands can be executed in a row when they are divided by a semicolon (;).

Example

XA2, 2, 0.5 ; XV4 ; XI2, AUTO CrLf

Note

• The total data length from the first character to the terminator should not exceed 200 bytes.

• A sub-delimita cannot be used for the commands FM, LF, CF, RF, RC, RS, BL, DS, XE, or XZ. These commands can only

be carried out one by one.

• In case of RS-232-C one ACK-status (E0/E1) will be output for each command devided by ;.