Sn sets the scaling unit, Xq sets the filter, Xv sets the measurement period – Yokogawa DR240 User Manual

Page 57: Xi sets the a/d integration time

Advertising
background image

6-3

IM DR231-11E

Commands

6

• For the measurement item p6, refer to the list of

measurement items on page 5-9 and 5-10.

• There are items that can not be set on the

measurement item, p6, depending on the wiring
method or the settings on other channels. Set
within the group indicated on page 5-10.

• Set the span values, p7 and p8, within the

measurement range indicated on page 5-8.

• Input values within 6 digits disregarding the

decimal point for p7 and p8. The decimal point
position is fixed to the decimal point position of
the setting range of page 5-8.

• For p9, p10, and p11, either set all the three

parameters or omit all three.

• The measurement range cannot be changed while

executing a report or computation, or when a file
is written.

Setting the strain input scaling(DR232/DR242)

Setting

SRp1, p2, p3, p4, p5, p6, p7, p8, p9<terminator>

p1

Channel number (001 to 560)

p2

SCL

p3

STRAIN

p4

Measurement range
2k/20k/200k

p5

Left value of span

p6

Right value of span

p7

Left value of scale
(-30000 to 30000)

p8

Right value of scale
(-30000 to 30000)

p9

Decimal point position of scale (0 to 4)

Example

Measure using the measurement range of 2k for
subunit 0 and channel 01, and convert 0

µε

to 0.00

and 1000

µε

to 100.00.

SR001, SCL, STRAIN, 2k, 0, 1000, 000, 10000, 2

Comments

• Even when the settings of p4 are same parameters,

the measurement range of the strain varies
depending on the gauge method. For details, refer
to page 5-8.

• Set spans p5 and p6 within the measurement range

indicated in page 5-8.

• Input values up to 6 digits disregarding the

decimal point for p5 and p6. The decimal point
position is fixed to the decimal point position of
the setting range of page 5-8.

• For p7, p9, and p10, either set all the three

parameters or omit all three.

• The measurement range cannot be changed while

executing a report or computation, or when a file
is written.

SN

Sets the scaling unit.

Mode

Operation mode

Setting

SNp1, p2<terminator>

p1

Channel number (001 to 560)
Computation channel number (A01 to A60)

p2

Unit character string (up to 6 characters)

Example

Set the scaling unit for channel 01 of subunit 0 to
ABCDEF.

SN001, ABCDEF

Comments

• For a character string that can be used as units, see

the characters in Section 5.6, “ASCII Code
Table.” However, note that a semi-colon “;”
cannot be used.

• The optional computation channel number for a

stand-alone model is A01 to A30.

XQ

Sets the filter.

Mode

Setup mode

Setting

XQp1<terminator>

p1

ON

Filter on

OFF

Filter off

Example

Insert a filter in the input.

XQON

XV

Sets the measurement period.

Mode

Setup mode

Setting

XVp1<terminator>

p1

Measurement interval (0.5, 1, 2, 3, 4, 5, 6,
10, 12, 15, 20, 30, or 60)

Example

Set the measurement period to 10 seconds.

XV10

Comments

• 0.5s and 1s are only for DR232 and DR242.
• The shortest measurement period that can be set

differs depending on the A/D integration time and
the filter on/off setting as follows.
Integration time: Equivalent to 50/60 Hz
Input module

Filter off

Filter on

10CH

0.5 s*

3 s

20CH

2 s

4 s

30CH

2 s

4 s

Integration time: 100 ms
Input module

Filter off

Filter on

10CH

4 s

12 s

20CH

5 s

15 s

30CH

6 s

20 s

*: 2s if the power monitor module is installed

XI

Sets the A/D integration time.

Mode

Setup mode

Setting

XIp1, p2<terminator>

p1

Subunit (0 to 5)

p2

Integration time (AUTO, 50 Hz, 60 Hz, 100
ms)

Example

Set the A/D integration time in subunit 0 to 50 Hz.

XI0, 50Hz

6.1 Setting the Input

Advertising