6-10

IM DR231-11E

6.5 Other Settings

SD

Sets the date and time.

Mode

Operation mode

Setting

SDp1, p2<terminator>

p1

Date (year, month, day)

p2

Time (hour, minute, second)

Example

Set the clock in the DR230/DR240 to 1 o’clock pm,
(0 min., 0 sec.) on July 1, 1996.

SD96/07/01, 13:00:00

Comments

• The formats for p1 and p2 are fixed at 8

characters. Set them in the following manner:
p1 = YY/MM/DD (Last two digits of the year,
month, day)
p2 = HH:MM:SS (Hour:minute:second)

• Do not place space(s) before and after, or embed

them in the parameter. Otherwise, an error occurs.

SV

Sets a moving average.

Mode

Operation mode

Setting

SVp1, p2<terminator>

p1

Channel number (001 to 560)

p2

Number of moving averages (0 to 64)
0 = Off

Example

Take moving averages 64 times in the input of
subunit 0, channel 01.

SV001, 64

SY

Sets how to copy the channel setting
parameters between channels.

Mode

Operation mode

Setting

SYp1, p2, p3, p4, p5-p6<terminator>

p1

Copy range parameters including unit (ON,
OFF)

p2

Copy alarm parameters (ON, OFF)

p3

Copy the other parameters (ON, OFF)

p4

Copy source channel numbers (001 to 560)
Computation channel number (A01 to A60)

p5

First channel number of copy destination
(001 to 560)
Computation channel number (A01 to A60)

p6

Last channel number of copy destination (01
to 60)

Example

Copy only the range setting data in channel 01 of
subunit 0, to channel 01 to channel 60 of subunit 1.

SYON, OFF, OFF, 001, 101-60

Copy only the range setting data in channel 02 of
subunit 0, to channel 03 of subunit 0.

SYON, OFF, OFF, 002, 003

Comments

• If the copy command is to be applied to successive

channels, insert a minus sign (-) between the first
channel number and the last channel number.
However, only channels in the same subunit can
be set successively.

• The optional computation channel number for a

stand-alone model is A01 to A30.

• No copying is possible between measurement and

computation channels.

SX

Sets channels in a group.

Mode

Operation mode

Setting

SXp1, p2<terminator>

p1

Group number (G01 to G07)

p2

Channel number (up to 36 characters)

Example

Set channels from channel 1 in slot 1 of unit 4 to
channel 10 in slot 3 of subunit 4 and channel 5 in slot
4 of unit 4 to group 3.

SXG03, 411-440, 445

Comments

• Delimit the channels set in p2 with a comma (,).

For successive channels, insert a “-” (minus)
between the first and last channels.

SI

Sets the timer.

Mode

Operation mode

Setting

SXp1, p2, p3, p4<terminator>

p1

Timer number (1 to 6)

p2

ABSOLUTE: Absolute time
RELATIVE: Relative time

p3

Time interval

For ABSOLUTE

1 min, 2 min, 3 min, 4 min, 5 min, 6 min, 10
min, 12 min, 15 min, 20 min, 30 min, 1 h, 2
h, 3 h, 4 h, 6 h, 8 h, 12 h, 24 h

For RELATIVE,

Fixed to 8 characters
DD HH:MM
DD: Day
HH: Hour
MM: Minute

p4

Reference time (fixed to 5 characters)
HH:MM
HH: Hour
MM: Minute

Example

Set timer No. 3 to intervals of 15 minutes each
starting at 15:15.

SI3, ABSOLUTE, 15 min, 15:15

Comments

• p4 is effective when p2 is ABSOLUTE.
• In ABSOLUTE, the time interval set with p3 is

used from the time set with p4.

• In RELATIVE, the time interval set with p3 is

used from the time when the timer setting is
completed, the power is turned on, or the set timer
is executed.

SQ

Sets a match time.

Mode

Operation mode

Setting

SQp1, p2<terminator>

p1

Match time number (1 to 3)

p2

Time (fixed to 11 characters)
DD HH:MM (day hour:minute)

Example

Set a match time 1 at 12:00 on the 15th day.

SQ1, 15 12:00:00

Comments

• When the time set here is reached, the operation

set for the match time is executed.

SL

Sets an event/action.

Mode

Operation mode

Release of event/action setting

Setting

SLp1, p2<terminator>

p1

event/action box number (01 to 30)

p2

NONE

Event action setting by an alarm or chart end event

Setting

SLp1, p2, p3, p4, p5<terminator>

p1

Event action box number (01 to 30)

p2

Event (ALARM, CHART_END)

p3

Action mode (EDGE, LEVL)

p4

Action
ALARM_ACK Acknowledgement of alarm

status.

ALARM_RST

Alarm reset

TIMER_RST

Timer reset

RECORD

Record start/stop

SPEED_CHG

Change to second chart
speed

MSG_PR

Message print

MANUAL_PR

Manual print

MSG_DISP

Display message

6.5 Other Settings