8 a/d calibration data output format – Yokogawa DR240 User Manual

Page 85

Advertising
background image

7-9

IM DR231-11E

Output Format

7

7.8

A/D Calibration Data Output Format

A/D calibration data are output in the following format by the TS8 + “Device Trigger (GET)” +
LF command: This is effective only in the A/D calibration mode.

XZ

A/D calibration data for the last slot

EN

Output completion

CrLf

CrLf

XZ

CrLf

A/D calibration data for the first slot

Calibration data are output for every slot after the command to execute A/D calibration. The
output format is the same form as in the XZ command setting shown below.

XZ subunit number, slot number, DISPLAY, calibration item CrLf

All calibration items are output.

Range of outputting

The number of slots to which A/D calibration data are output is determined by the channels
specified by the LF command. For example, if channels 042 to 236 are specified by the LF
command, data from subunit 0, slot 4 to subunit 2, slot 3 are output.