IM DR231-11E

Commands

6

6.7 Data Output Request

Command

TS

Selects the output data.

Mode

All modes (A/D calibration data output is limited to A/
D calibration mode only.)

Setting

TSp1<terminator>

p1

0 Measured data output
1 Setting data output
2 Unit data output
4 Report data output
5 System configuration data output
8 A/D calibration data output
9 Setting data output in setup mode

Comments

• The setting for p1=0, 1 or 2 is effective only in the

operation mode.

• The setting for p1=8 is effective only in the A/D

calibration mode.

• The setting “p1=3” is valid only for recorders with

a floppy disk drive function .

• Do not output data from the RAM disk using a

TS3 command during computation.

• The setting, p1=4 is effective when the instrument

has the report function and one of hourly, daily
and monthly reports is ON.

FM

Selects the output format for
measured/ computed data.

Mode

Operation mode

Setting

FMp1, p2, p3<terminator>

p1

0 Outputs measured data in ASCII format.
1 Outputs measured data in binary format.
2 Outputs computed data in ASCII format.
3 Outputs computed data in binary format.

p2

First output channel (001 to 560)
First output computed channel (A01 to A60)

p3

Last output channel (001 to 560)
Last output computed channel (A01 to A60)

Comments

• You must specify data to be output with the TS

command and execute “GET” or “ESC T”before
sending the FM command.

• The command selects the output format of

measured/computed data, either ASCII or binary,
and the output channel.

• If no input channel is recognized by the DR130/

DR230/DR240 among the specified channels, a
syntax error occurs.

• The setting “p1=2” or “p1=3” is valid only for

recorders with the optional computing function or
floppy disk drive function or with pulse modules
installed.

• The optional computation channel number for a

stand-alone model is A01 to A30.

MF

Sets the output formats of the
measured data/computed data on a
RAM disk.

Mode

Operation mode

Setting

MFp1, p2, p3, p4, p5, p6<terminator>

p1

Output contents

0

File directory

1

Measured/computed data output in
ASCII format.

2

Measured/computed data output in
binary format

3

On/off information on channel on which
specified file exists

p2

File name (up to eight characters)

p3

First output channel number

p4

Last output channel number

p5

First output data number

p6

Last output data number

Comments

• This command is effective only with an optional

floppy disk drive.

• Always specify data that are output with TS

command before sending MF command, and
execute [GET] or [ESCT].

• If p1 is set to 0 (file directory), the settings for p2

to p6 will be invalid.

• If p1 is set to 3 (on/off information on the channel

on which the specified file
exists), the settings for p5 and p6 will be invalid.

• Parameters p5 and p6 are set in the range from 1

to the last data number in a file.

• Channel numbers are recognized in the order of

input channel and channel for computation. If you
set 005 for the first channel and A10 for the last
channel, the target computation will be done in
input channel 005 to the last input channel as well
as channels A01 to A10.

RF

Selects the output format of the
report.

Mode

Operation mode

Setting

RFp1, p2, p3<terminator>

p1

0 Output hourly report data
1 Output daily report data
2 Output monthly report data
3 Output the status of the hourly/

daily/monthly data

p2

First channel for output (R01 to R60)

p3

Last channel for output (R01 to R60)

Comments

• Before transmitting the RF command, be sure to

specify the data to output using the TS command
and execute “GET” or “ESC T.”

• If the daily report is to be output using extended

format, the extended information must be output
within 1 hour from the creation of the report. If
the monthly report is to be output using extended
format, the extended information must be output
within 1 day from the creation of the report. After
that, the extended information can not be output.

• If there is no valid data within the specified report

channel range, “FFFFH” is output.

LF

Specifies the output channels for
setting data output, unit, and decimal
point data.

Mode

All modes

Setting

LFp1, p2<terminator>

p1

First output channel (001 to 560)
First output computed channel (A01 to A60)

p2

Last output channel (001 to 560)
Last output computedchannel (A01 to A60)

Comments

• You must specify data to be output with the TS

command and execute “GET” or “ESC T” before
sending the LF command.

• If no input channel is recognized by the DR230/

DR240 among the specified channels, a syntax
error occurs.

• The optional computation channel number for a

stand-alone model is A01 to A30.

6.7 Data Output Request Command