2 command syntax – Yokogawa DR240 User Manual

Page 46

Advertising
background image

5-3

IM DR231-11E

Command Format

5

5.2

Command Syntax

In this manual, each command is explained as shown below.

SD Set date & time.

Mode

Operation mode

Set

SDp1, p2<terminator>

p1 year, month, day
p2 hour, min., sec.

Ex.

Set the clock inside the DA to July 1, ’96, 13:00:00

SD96/07/01, 13:00:00

Comments

• The formats of p1 and p2 are fixed to 8 characters.

Set them in the following manner.
p1=YY/MM/DD
p2=HH:MM:SS

Command function
Command

Effective command mode
Command syntax

Parameter description

Example of setting

Explanation of the way to
use and remarks in more detail

Mode

There are the following three modes in the DR130/DR230/DR240.
• Operation (measurement) mode

Mode to perform normal operation (measurement).

• Setup mode

Mode to set the basic specifications for the DR130/DR230/DR240.
To confirm the settings, execute the XESTORE command. See the description for the XE command.

• A/D calibration mode

Mode to execute calibration of the A/D module.

Each mode is selected by the DS command. For details, see DS Command on page 6-20.
Also in each mode, commands and parameters that can be set and parameters that can be output
differ. See the description for each command.

Number of Channels

The configurable numbers of measurement and computation channels vary, from recorder to recorder,
between model DR130, stand-alone models of the DR230/240, and expandable models of the DR230/
240, as shown below:

Type of DR Recorder Highest Measurement-channel

Highest Computation-channel

Number

Number

DR130

020

A30

Stand-alone models

030

A30

of DR230/240
Expandable models

560

A60

of DR230/240

Advertising