7 transferring the instantaneous v alues – Yokogawa DR240 User Manual

Page 39

Advertising
background image

4-10

IM DR231-11E

4.7

Transferring the Instantaneous V alues

The instantaneous values of the data measured on the DR (current measured data) are transmitted to
the PC that is connected through port No. 34151.

Note

• Up to four PCs can connect to port No. 34151 of one DR.

• The commands described here do not affect the status byte.

• EF, EL, and EB command do not support sub-delimiters.

The measured data are transmitted using the following command.

EF

Outputs the measured/computed data in binary format.

Mode

Operation mode

Syntax

EFp1, p2, p3<terminator>

p1

Data to be output

0 Output only the measured/computed data.
1 Output the measured/computed data and alarm data

p2

First channel to output (001 to 560, or A01 to A60)

p3

Last channel to output (001 to 560, or A01 to A60)

Description

• A01 to A60 in p2 and p3 correspond to computation channels (A01 to A30 for the

stand-alone type).

• The data of expansion type channels that are not connected are not output.
• The data are output in the byte order specified by the EB command.
• Measured data and computed data are output simultaneously.
• If a parameter is omitted, the parameter specified previously is used.
• If a specified channel cannot output data, 2-byte data with a data length of zero

are output.

• The output format is as follows.

Year

Month

Day

Hour

Min

Sec

MS

DM

Data length

A1

B1

C1

D1

Measured data

Computed data

E1

F1

G1

H1

Am

Bm

Cm

Dm

Measured data

Computed data

En

Fn

Gn

Hn

Data length:

When p1=0: 8 + measurement ch

×

4 + computation ch

×

6

When p1=1: 8 + measurement ch

×

6 + computation ch

×

8

MS: Value in units of 0.1 s. 0 or 5.
DM: Dummy (undefined).
A1 to An: Unit number. Computation channel is fixed to 0

×

80.

B1 to Bn: Measurement channel number
C1 to Cn, G1 to Gn: Alarm status (level 1, 2)

(No output when p1 is 0)
0: No warning
1: Upper limit alarm
2: Lower limit alarm
3: Upper difference limit alarm
4: Lower difference limit alarm
5: Rate-of-change upper-limit alarm
6: Rate-of-change lower limit alarm