6 rs-232-c interface parameter setting procedure – Yokogawa DR240 User Manual

Page 19

Advertising
background image

2-9

IM DR231-11E

Overview and Specifications of RS-232-C Interface

2

2.6

RS-232-C Interface Parameter Setting Procedure

Setting of the RS-232-C parameters must be carried out using the 3 dipswitches located next to the
module connector.

1

2

3

4

Parameter setting
switch 1

Data length

Baud rate

1

2

3

4

Parameter setting
switch 2

Baud rate

Parity

Stop bit

1

2

3

4

Parameter setting
switch 3

Not use

Handshake system

ON

OFF

OFF

OFF

ON

ON

Baud rate (No.1 to 3 of setting switch 1 and No.4 of setting switch 2)

Baud rate

No.1

No.2

No.3

No.4 (switch 2)

150

OFF

OFF

OFF

OFF

300

OFF

OFF

ON

OFF

600

OFF

ON

OFF

OFF

1200

OFF

ON

ON

OFF

2400

ON

OFF

OFF

OFF

4800

ON

OFF

ON

OFF

9600

ON

ON

OFF

OFF

Default

19200

ON

ON

ON

OFF

38400

OFF

OFF

OFF

ON

Data length (Switch No.4 of parameter setting switch 1)

Data length

No.4

7

OFF

8

ON

Default

Parity (Switch No.1 and 2 of parameter setting switch 2)

Parity

No.1

No.2

NONE

OFF

OFF

ODD

OFF

ON

EVEN

ON

OFF

Default

Stop bit (Switch No.3 of parameter setting switch 2)

Stop bit

No.4

1

OFF

Default

2

ON

Handshake system (Switch No.1 to 3 of parameter setting switch 3)

Handshake system

No.1

No.2

No.3

No handshake

OFF

OFF

OFF

Default

XON-ER*

OFF

OFF

ON

XON-RS*

OFF

ON

OFF

CS-ER

OFF

ON

ON

CS-RS

ON

OFF

OFF

* When the baud rate is set to 38400, there is no handshaking.