6-2

IM DR231-11E

• Set the span values, p6 and p7, within the

measurement range indicated on page 5-8.

• Input values within 6 digits disregarding the

decimal point for p6 and p7. The decimal point
position is fixed to the decimal point position of
the setting range of page 5-8.

• The measurement range cannot be changed while

executing a report.

• The measurement range cannot be changed while

executing a report or computation, or when a file
is written.

Setting the strain input(DR232/DR242)

Setting

SRp1, p2, p3, p4, p5<terminator>

p1

Channel number (001 to 560)

p2

STRAIN

p3

Measurement range
2k/20k/200k

p4

Left value of span

p5

Right value of span

Example

Set the measurement range of subunit 1 and channel
01 to 20k, left value of span to 0, and right value of
span to 100.

SR101, STRAIN, 20k, 0, 100

Comments

• Even when the settings of p3 are the same

parameters, the measurement range of the strain
varies depending on the gauge method. For
details, refer to page 5-9.

• Set spans, p4 and p5, within the measurement

range indicated in page 5-9.

• Input values within 6 digits disregarding the

decimal point for p4 and p5.

• The measurement range cannot be changed while

executing a report or compuration, or when a file
is written.

Setting the pulse input(DR232/DR242)

Setting

SRp1, p2, p3, p4, p5, p6, p7, p8, p9<terminator>

p1

Channel number (001 to 560)

p2

PULSE

p3

Measuring mode
RATE momentary pulse count mode
GATE contact ON/OFF detection mode

p4

Left value of span (0 to 30000)

p5

Right value of span (0 to 30000)

p6

Left value of scale (0 to 30000)

p7

Right value of scale (0 to 30000)

p8

Decimal point position of scale (0 to 4)

p9

Filter (ON or OFF)

Example

Set subunit 1 and channel 01 to momentary pulse
count mode, left value of span to 0, and right value
of span to 1000.

SR101, PULSE, RATE, 0, 1000, 0, 1000, 0

Comments

• When omitting p6, p7, or p8, omit all three.
• When summing the count number or the ON time,

set the computational expression with the SO
command, then execute computation start with the
EX command. When the pulse input module is
installed, the use of computation channels become
available without the math option. However, the
only computational expression available in this
case is TLOG.PSUM(XXX) for summing the
count number or the ON time.

• The measurement range cannot be changed while

executing a report or computation, or when a file
is written.

Setting Scaling

Setting

SRp1, p2, p3, p4, p5, p6, p7, p8, p9<terminator>

p1

Channel number (001 to 560)
Conputation channel number (A01 to A60)

p2

SCL

p3

Type of input
VOLT DC voltage
TC

Thermocouple

RTD

Resistance temperature detector

DI

Contact

mA

DC current

p4

Measurement range

p5

Left value of span

p6

Right value of span

p7

Left value of scale (-30000 to 30000)

p8

Right value of scale (-30000 to 30000)

p9

Decimal point position of scale (0 to 4)

Example

Change channel 02 of subunit 0 to an input in which
1 V is converted to 0.00 and 5 V to 100.00.

SR002, SCL, VOLT, 6V, 1000, 5000, 0, 10000, 2

Comments

• For the p4 measurement range, see the range

parameter table on Pages 5-7 and 5-8.

• For the p5 and p6 setting spans, set them in the

ranges shown in the setting range on Pages 5-7
and 5-8.

• For p5 and p6 setting values, enter them within 6

digits excluding the decimal point.

• Select either to set all three parameters, p7, p8 and

p9, or to omit them all.

• The measurement range cannot be changed while

executing a report or computation, or when a file
is written.

Setting the power monitor scaling

Setting SRp1, p2, p3, p4, p5, p6, p7, p8, p9, p10, p11<terminator>

p1

Channel number (001 to 560)

p2

SCL

p3

AC

p4

Wiring method

1Ph2W:

Single-phase two-wire system

1Ph3W:

Single-phase three-wire system
(only on a three-wire system)

3Ph3W-2I: Three-phase three-wire system

(only on a two-voltage two-
current three-wire system)

3Ph3W-3I: Three-phase three-wire system

(only on a three-voltage three-
current three-wire system)

3Ph4W:

three-phase four-wire system
(only on a three-wire system)

p5

Input range
25 V - 0.5 A
25 V - 5 A
250 V - 0.5 A
250 V - 5 A

p6

Measurement item

p7

Left value of span

p8

Right value of span

p9

Left value of scale (-30000 to 30000)

p10

Right value of scale (-30000 to 30000)

p11

Decimal point position of scale (0 to 4)

Example

Measure the active power on subunit 0 and channel
01, and convert 10W to 0.00 and 100W to 100.00.

SR001, SCL, AC, 3Ph3W-2I, 250V-0.5A, P1, 10,
100, 000, 10000, 2

Comments

• Settings of p4 and p5 are common within one

power monitor module. If the setting is changed,
the settings on other channels also change.

6.1 Setting the Input