Pl turns on or off list printing for every channel, Xr performs settings related to records, Xc sets dot printing colors – Yokogawa DR240 User Manual
Page 63

6-9
IM DR231-11E
Commands
6
• If ON2 is set to a channel whose recording width
is less than 100 mm, the setting is invalid.
Printing is carried out at positions 0% and 100%
only.
• The optional computation channel number for a
stand-alone model is A01 to A30.
PL
Turns on or off list printing for every
channel.
Mode
Operation mode
Setting
PLp1, p2<terminator>
p1
Channel number (001 to 560)
Computation channel number (A01 to A60)
p2
ON/OFF
Example
Perform list printing for channel 6 in slot 3 of subunit
4.
PL436, ON
Comments
• The optional computation channel number for a
stand-alone model is A01 to A30.
XR
Performs settings related to records.
Mode
Setup mode
Setting
XRp1, p2, p3, p4, p5, p6, p7, p8, p9, p10,
p11<terminator>
p1
CHANNEL Channel number recording/
displaying
TAG
Tag recording/displaying
p2
Number of characters recorded in a tag (7 to
16)
p3
Analog trend recording mode
NORMAL
Normal analog trend recording
ALARM1
Analog trend recording starts
at an alarm occurrence.
ALARM2
Analog trend recording starts
at an alarm occurrence and
stops at an alarm reset.
GROUP
Analog trend recording of a
group set for an event action.
p4
Turning on/off of chart speed change
printing (ON, OFF)
p5
Turning on/off of printing “PRINT ON”
(ON, OFF)
p6
Number of channels to be recorded in a line
for time printing (1 to 4, only 1 or 2 for the
DR130)
p7
Channel printing interval
OFF
No printing
5.0 mm
5 mm
12.5 mm
12.5 mm
p8
Title printing interval
OFF
No printing
600 mm
600 mm
1500 mm
1500 mm
p9
Turning on/off of scale check (ON, OFF)
p10
Logging record format
HORIZON
Horizontal writing
VERTICAL Vertical writing
p11
Selection of printing interval in logging or
digital recording
SINGLE
Logging printing is recorded at
the time interval set in timer
No. 1.
In digital printing, the printing
interval is automatically
determined by the chart speed.
MULTIPLE Recording is done at the
recording interval set for each
channel.
Example
Perform tag display/recording with 7 characters,
analog recording when an alarm occurs, time printing
in 4 channels in a line, channel printing interval of
12.5 mm, title printing with 600 mm interval, and
horizontal writing of logging records and at the
intervals set for each channel.
XRTAG, 7, ALARM1, OFF, OFF, 4, 12.5 mm, 600
mm, OFF, HORIZON, MULTIPLE
Comments
• Set the timer with the SI command (see 6-10).
• If p11 is set to MULTIPLE, recording is done at
the interval set with the PD command.
• If p11 is set to SINGLE, the digital printout
recording interval isautomatically determined
according to the chart speed, and the number
ofchannels and rows in which numeric values are
printed. For details, seethe separate manual (IM
DR231-01E or IM DR232-01E).
XC
Sets dot printing colors.
Mode
Setup mode
Setting
XCp1, p2<terminator>
p1
Channel number (001 to 560)
Computation channel number (A01 to A60)
p2
Dot printing color
PURPLE
RED
GREEN
BLUE
BROWN
BLACK
NAVY(Navy blue)
YEL-GR(Yellow-green)
RED-PR(Red-purple)
ORANGE
Example
Record channel 6 in slot 3 of subunit 4 in red-purple.
XC436, RED-PR
Comments
• The optional computation channel number for a
stand-alone model is A01 to A30.
6.4 Setting the Recording