Example

Calibrate the 20-mV range on the module in slot 1 of
subunit 0.

XZ0, 1, CAL/EXEC, 20 mV

Correcting the calibration value manually
(universal module)

Setting

XZp1, p2, p3, ..... , p25<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 20-mV range

p5

Span calibration value for 20-mV range

p6

Zero calibration value for 60-mV range

p7

Span calibration value for 60-mV range

p8

Zero calibration value for 200-mV range

p9

Span calibration value for 200-mV range

p10

Zero calibration value for 2-V range

p11

Span calibration value for 2-V range

p12

Zero calibration value for 6-V range

p13

Span calibration value for 6-V range

p14

Zero calibration value for 20-V range

p15

Span calibration value for 20-V range

p16

Zero calibration value for 50-V range

p17

Span calibration value for 50-V range

p18

Zero calibration value for Pt:1 mA

p19

Span calibration value for Pt:1 mA

p20

Zero calibration value for Pt:2 mA

p21

Span calibration value for Pt:2 mA

p22

Span calibration value for Pt:1 mA-H

p23

Span calibration value for Pt:2 mA-H

p24

Zero calibration value for Cu:2 mA

p25

Span calibration value for Cu:2 mA

Correcting the calibration value manually (DCV/TC/
DI input module)

Setting

XZp1, p2, ..... , p17<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 20-mV range

p5

Span calibration value for 20-mV range

p6

Zero calibration value for 60-mV range

p7

Span calibration value for 60-mV range

p8

Zero calibration value for 200-mV range

p9

Span calibration value for 200-mV range

p10

Zero calibration value for 2-V range

p11

Span calibration value for 2-V range

p12

Zero calibration value for 6-V range

p13

Span calibration value for 6-V range

p14

Zero calibration value for 20-V range

p15

Span calibration value for 20-V range

p16

Zero calibration value for 50-V range

p17

Span calibration value for 50-V range

Correcting the calibration value manually (mA
input module)

Setting

XZp1, p2, p3, p4, p5<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 20-mA range

p5

Span calibration value for 20-mA range

Correcting the calibration value manually (Strain
input module)

Setting

XZp1, p2, p3, ....., p9<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 2k range

p5

Span calibration value for 2k range

p6

Zero calibration value for 20k range

p7

Span calibration value for 20k range

p8

Zero calibration value for 200k range

p9

Span calibration value for 200k range

Correcting the calibration value manually (Digital
input module)

Setting

XZp1, p2, ..... , p7<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

DISPLAY

p4

Zero calibration value for 60-mV range

p5

Span calibration value for 60-mV range

p6

Zero calibration value for 6-V range

p7

Span calibration value for 6-V range

Example

Modify the span calibration value for the 6V range of
the module in slot 1 of subunit 0 to 32000.

XZ0, 1, DISPLAY, , , , 32000

Storage of A/D calibration data in a module (A/D
calibration END processing)

Setting

XZp1, p2, p3, p4<terminator>

p1

Unit number (0 to 5)

p2

Slot number (0 to 5)

p3

END

p4

Storage selection (ABORT, STORE)

Example

Store the calibrated values for the module in slot 1 of
subunit 0.

XZ0, 1, END, STORE

Comments

• You must execute STORE using A/D calibration

end processing each time after making an A/D
calibration or adjustment for one module. If A/D
calibration end processing (STORE) is not
executed, the calibrated data will not take effect.

• Since it takes an indefinite time to execute the XZ

command, return an ACK after the processing
ends.

• On the controller side, execute the following

processing after receiving an ACK after
transmitting the XZ command. The format of the
ACK status is as follows:
E0Cr+Lf

The received command was normally
processed.

E1Cr+Lf

There is an error in the received
command.

• If A/D calibration data are adjusted, inputting a

numeric value which is out of the setting range
disables the modules to be recognized. For
details, see the separate manual (IM DR231-01E
or IM DR232-01E).