Fv stores setting data on a floppy disk, Fl reads the setting data from a floppy disk, Fe deletes a file on a floppy disk of setting data – Yokogawa DR240 User Manual

Page 73: Ye deletes a file on a floppy disk(setup mode), Dr start/stop the report

Advertising
background image

6-19

IM DR231-11E

Commands

6

• Parameters p7 and p8 are set in the range from 1

to the last data number in the file.

• Channel numbers are recognized in the order of

input channel and channel for computation. If you
set 005 for the first channel and A10 for the last
channel, an ASCII conversion will be made in
input channel 005 to the last input channel as well
as channels A01 to A10 for computation.

• This command is not executable during

computation.

FV

Stores setting data on a floppy disk.

Mode

Operation mode

Setting

FVp1<terminator>

p1

File name (up to eight characters)

Example

Store the setting data with the file name SET1.

FVSET1

Comments

• This command is effective only with an optional

floppy disk drive function.

• Setting data in the setup mode cannot be stored.

To store them, use a YV command.

• This command is not executable during

computation.

FL

Reads the setting data from a floppy
disk.

Mode

Operation mode

Setting

FLp1 p2<terminator>

p1

Method of reading data:

DIRECT

Starts reading data
immediately after sending an
FL command.

TRIG_1 to 3 Starts reading data at the same

time as an event occurrence in
event/action functions.

p2

File name

Example

Read setting data with the file name SET1
immediately.

FLDIRECT, SET1

Comments

• This command is effective only with an optional

floppy disk drive function.

• Setting data in the setup mode cannot be read. To

read them, use a YL command.

• If TRIG_1 to TRIG_3 are set to p1, event/action

functions need to set FLOPY:LD_TRG1 to 3 for
their actions.

• This command is not executable during

computation.

FE

Deletes a file on a floppy disk of
setting data.

Mode

Operation mode

Setting

FEp1<terminator>

p1

File name

Example

Delete the file SET2.

FESET2

Comments

• This command is effective only with an optional

floppy disk drive function.

• A data file in the setup mode cannot be deleted.

To delete it, use a YE command.

YV

Stores set data in the setup mode on
a floppy disk.

Mode

Setup mode

Setting

YVp1<terminator>

p1

File name (up to eight characters)

Example

Store set data in the setup mode with the file name
SET1.

YVSET1

Comments

• This command is effective only with an optional

floppy disk drive function.

YL

Reads set data in the setup mode
from a floppy disk.

Mode

Setup mode

Setting

YLp1<terminator>

p1

File name (up to eight characters)

Example

Read set data in the setup mode, which are on the
disk with the file name SET1.

YLSET1

Comments

• This command is effective only with an optional

floppy disk drive function.

YE

Deletes a file on a floppy disk(Setup
Mode).

Mode

Setup mode

Setting

YEp1<terminator>

p1

File name

Example

Delete the file SET2.

YESET2

Comments

• This command is effective only with an optional

floppy disk drive function.

BL

Executes the initial balancing of the
strain input channel

Mode

Operation mode

Setting

BLp1, p2, p3<terminator>

p1

The first channel for executing the initial
balancing

p2

The last channel for executing the initial
balancing

p3

Select either initial balancing or initialization

EXEC: Execute initial balancing
INIT: Execute initialization

Example

Execute initial balancing on subunit 0 and channels
01 to 08.

BL001, 008, EXEC

Comments

• Channels other than strain input channels or

channels that are not connected within the
specified range are ignored.

• Initial balancing and initialization can not be

executed during a report.

• If initial balancing is executed, number-of-

channels worth of data are returned in the
following format.
S1 S2 CCC DDCrLf

S1:

Data status 1
N: Normal
S: SKIP

S2:

Data status 2
Space: Data in the middle
E: Data at the end

CCC: Channel number
DD: Result of the initial balancing

OK: Initial balancing succeeded
NG: Initial balancing failed
DF: Default values set
_: Skip module

DR

Start/stop the report

Mode

Operation mode

Setting

DRp1<terminator>

p1

Start/stop the report

6.6 Control and Execution Command

Advertising