Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 19

CHAPTER 3

INSTALLATION

This chapter provides you with information that is intended to help you

install the GMW90 series transmitters.

Configuration Before Installation

If you need to change the settings of the transmitter, it is best to do this

before it has been installed. Available configuration options are different

for analog output models (such as GMW93) and digital output models

(for example, GMW95).

Configuration of Analog Output Models

You can configure analog output models of GMW90 series transmitters

in two ways:

- Using the DIP switches on the component board. See section DIP

Switches of Analog Output Models on page 20.

- Configuring the settings in software through the service port. See

connection instructions and serial line commands in Chapter 4,

Operation, on page 31.

These two configuration methods are mutually exclusive. If the DIP

switch configuration is used, software settings have no effect on settings

that are controlled by the DIP switches. Additionally, standard analog

output settings and display configuration are always set in the DIP

configuration mode. DIP switch number 8 is the master switch that

controls which configuration method is used.