Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 53

Calibration and Adjustment Commands

The following sections describe the calibration and adjustment

commands of the GMW90 series. For general information on performing

calibration and adjustment on the serial line, see section Adjustment

Using a Computer on page 68.

NOTE

You must enable advanced serial commands using the PASS 9000

command before using the calibration and adjustment commands.

Adjust CO

2

Measurement

Use the CCO2 command to adjust the carbon dioxide (CO

2

)

measurement. You can do a 1-point or a 2-point adjustment, or clear the

adjustment information from the GM10 module. Note that the factory

adjustment will remain intact when user adjustment is cleared.

Remember to let the transmitter stabilize in the reference concentration

before entering the adjustment command. Three minutes is typically

enough.

NOTE

Before using the CCO2 command, read section Notes for CO2

Adjustment on page 64.

Show Current CO

2

Adjustment

In addition to the user-adjustable offset and gain values, the CCO2

command displays diagnostic information that may be useful to Vaisala

Helpdesk if there is a problem with the adjustment.

CCO2<cr>

Example:

>pass 9000

>cco2

User gain : 1.000

User offset : -8.365

CO2 (pre-adjust) : 464.625

30s delta : 13.904

30s stddev : 4.429

Adjustability : CAN_ADJUST