Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 71

Replacing the HTM10 Module

CAUTION

Handle the HTM10 module carefully. When reinstalling the transmitter

body to the mounting base, avoid touching the module or the

HUMICAP

®

sensor.

To replace the module:
1.

Disconnect the transmitter body from the mounting base.

2.

With your fingers, push apart the two plastic holders that hold the

module. Pull out the module. Keep the module straight while

pulling it out, otherwise the pins may twist in the connector and

damage it.

1112-028

Figure 29

Replacing the HTM10 Module

3.

Take the new module and align the pins to the connector on the

transmitter’s component board.

4.

Push down on the module so that the plastic holders clip into place.

Try not to tilt the module, so that the pins will meet the connector

straight on.

5.

Reconnect the transmitter to the mounting base.

6.

Verify that there are no errors when the transmitter starts up. If

there are, see section Error Messages on page 73.