74 ___________________________________________________________________ M211659EN-B

Viewing Error Messages on Serial Line

View Currently Active Errors

Use the ERRS command to view currently active errors on the serial

line:

ERRS<cr>

Example:

>errs

NO ERRORS

View Error Table

Use the ERRT command to view the table of possible transmitter errors.

The table includes error ID, error count since last reset, level, current

state, and error text.

Critical errors require a transmitter reset to recover. Other errors may be

recoverable if their cause is removed.

ERRT<cr>

Example:

>errt

Id: N: Level:State: Error text

1: 0: CRITICAL:OFF: FLASH memory corrupted

2: 0: CRITICAL:OFF: Parameter read (using defaults)

3: 0: CRITICAL:OFF: Parameter write

4: 0: CRITICAL:OFF: HTM10 03 FLASH Corrupted

21: 0: ERROR:OFF: HTM10 04 RH measurement

22: 0: ERROR:OFF: HTM10 05 T measurement

23: 0: ERROR:OFF: HTM10 01 Continuous communication failure

42: 0: ERROR:OFF: HTM10 06 Device Descriptor mismatch

43: 0: CRITICAL:OFF: Factory parameter memory not consistent

79: 0: ERROR:OFF: GM10 01 Continuous communication failure

81: 0: CRITICAL:OFF: GM10 03 FLASH Corrupted

82: 0: CRITICAL:OFF: GM10 04 MEM I

28: 0: ERROR:OFF: GM10 10 Device Descriptor mismatch

88: 0: ERROR:OFF: GM10 11 Module uncalibrated

89: 0: ERROR:OFF: GM10 12 CO2 measurement