Show command list, Measurement settings, Set environmental parameters – Vaisala GMW90 User Manual

Page 44

Advertising
background image

User's Guide _______________________________________________________________________

42 ___________________________________________________________________ M211659EN-B

Show Command List

Use the HELP command to list the currently available serial commands.

If the PASS command has not been used, only the basic serial commands

are available.

HELP<cr>

Example (basic serial commands from a transmitter model with analog

outputs, advanced commands are not enabled):

>help

?

CALCS

ECHO

ENV

ERRT

ERRS

FORM

HELP

INTV

PASS

R

RESET

SDELAY

SEND

SNUM

STATUS

UNIT

VERS

Measurement Settings

Set Environmental Parameters

Use the ENV command to set environmental parameters that affect the

measurement. For GMW90 series transmitters you can set the ambient

pressure value that is used for pressure compensation of CO

2

measurement and calculated humidity parameters.

ENV [pressure]<cr>

where
pressure = Ambient pressure in hPa.

Example:

>env 1013.3

Pressure (hPa) : 1013.3