Display settings, Select parameters to display – Vaisala GMW90 User Manual

Page 49

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 47

Display Settings

Select Parameters to Display

Use the DSEL command to select the parameters that are displayed on

the transmitter screen. You can select parameters by abbreviation. If only

one parameter is selected, it is shown vertically centered on the

transmitter screen.
DSEL [Q1 Q2 Q3]<cr>

where
Q1 = First parameter to show on the screen. Available parameters

are:

out1

same parameter as analog output channel 1

out2

same parameter as analog output channel 2

out3

same parameter as analog output channel 3

CO2

carbon dioxide level

CO2_10

carbon dioxide level in 10 ppm steps

RH

relative humidity

T

temperature

Tdf

dew/frostpoint temperature

Td

dewpoint temperature

Tw

wetbulb temperature

h

enthalpy

x

mixing ratio

a

absolute humidity

dTd

dew/frostpoint depression

Q2 = Second parameter to show on the screen. Available

parameters are the same as for Q1.

Q3 = Third parameter to show on the screen. Available parameters

are the same as for Q1.

Example (show currently displayed parameters):

>pass 9000

>dsel

Quant 1 : RH

Quant 2 : T

Quant 3 : CO2

Example (change display to only show CO2):

>dsel CO2

OK

Advertising