Reverting to factory settings using service port – Vaisala GMW90 User Manual

Page 79

Advertising
background image

Chapter 6 ____________________________________________________________ Troubleshooting

VAISALA ________________________________________________________________________ 77

4.

Reconnect the transmitter cover to the mounting base so the

transmitter powers up. Check the screen after power-up: when the

DIP switches are in factory reset position, you will see a

notification text.

5.

Disconnect the transmitter cover again.

6.

Set the DIP switches to the positions they were before. If you are

restoring the factory configuration on a factory customized analog

output model (GMW90 or GMW90R), leave DIP number 8 to the

Custom position to avoid the custom configuration from being

overwritted by the default settings at transmitter startup. For more

information about this, see section Changing Between DIP and

Custom Configuration on page 21.

7.

Reconnect the transmitter cover to the mounting base. Check the

startup screens to verify the configuration.

Reverting to Factory Settings Using

Service Port

Use the FRESTORE command to restore the transmitter to factory

settings.

FRESTORE<cr>

NOTE

If you are restoring the factory configuration on a factory customized

analog output model (GMW90 or GMW90R), leave DIP number 8 to the

Custom position to avoid the custom configuration from being

overwritted by the default settings at transmitter reset. For more

information about this, see section Changing Between DIP and Custom

Configuration on page 21.

NOTE

After using the FRESTORE command, reset the transmitter using the

RESET command.

Example:

>pass 9000

>frestore

Restoring HTM10 factory parameters

HTM10 factory parameters restored

Restoring GM10 factory parameters

GM10 factory parameters restored

Restoring GMW95R factory settings

110/110 parameters restored

>reset

Resetting

GMW95R / 1.1.28.5849 / XM90

>

Advertising