Vaisala GMW90 User Manual

Page 48

Advertising
background image

User's Guide _______________________________________________________________________

46 ___________________________________________________________________ M211659EN-B

Now give the following AOVER command:

>aover 3 5 10

Aout 3 clipping : 5.00 %

Aout 3 error limit : 10.00 %

Channel 3 now behaves like this:

- Clipping is now set to 5%, meaning the output is allowed to vary

between 0 ... 5.25 V. The channel will output the measurement for

0 ... 2100 ppm, but range 0 ... 5 V remains scaled to 0 ... 2000 ppm.

- Error limit is 10%, which means channel 3 will show the error state

(5.5 V) when the measured value is 10% outside the scaled output

range. With the settings above, this will happen if the measured

temperature is outside range 0 ... 2100 ppm.

- The output will never actually be between 5.25 and 5.5 V because of

clipping.

Advertising