Troubleshooting, Problem situations, Chapter 6 – Vaisala GMW90 User Manual

Page 74: Chapter 6, troubleshooting

Advertising
background image

User's Guide _______________________________________________________________________

72 ___________________________________________________________________ M211659EN-B

CHAPTER 6

TROUBLESHOOTING

This chapter describes possible problems, their probable causes and

remedies, and provides contact information for technical support.

Problem Situations

Table 11

Troubleshooting Table

Problem

Possible Cause

Remedy

Transmitter reports an

error on the display or

serial line.

Multiple causes.

Refer to Table 12 on

page 73.

Temperature reading

shown by the transmitter

is too high.

Transmitter is installed in

an unsuitable location,

for example, near a heat

source or in sunlight.

Relocate transmitter.

See section Selecting

Location on page 24.

Transmitter installed in

improper orientation.

Reinstall transmitter with

the arrow on the

mounting base pointing

up.

Transmitter does not

recognize a valid serial

command, responds with

message

FAIL 1: Unknown

command

Advanced commands

not enabled.

Issue the

PASS 9000

command.

Remote echo enabled on

the transmitter, causing

collisions on the RS-485

line.

Issue the

ECHO OFF

command to disable

remote echo.

Intermittent connection

problem between the

transmitter and your

terminal.

Issue the command

again.

Cannot see what you are

typing on the serial line.

Serial line echo not

enabled.

Enable local echo on

your terminal. See

section Terminal

Application Settings on

page 35.

Unable to complete CO

2

adjustment successfully.

Incorrect reference gas

flow.

Retry calibration after

reading section

Calibration and

Adjustment on page 63.

Breathing on transmitter

while using ambient gas

as reference.

Measurement not

stabilized before

attempting adjustment.

Incorrect calibration gas

concentration(s) used for

2-point calibration.