Adjustment using a computer, Adjustment, Using a computer – Vaisala GMW90 User Manual

Page 70

Advertising
background image

User's Guide _______________________________________________________________________

68 ___________________________________________________________________ M211659EN-B

Adjustment Using a Computer

For more detailed instructions on using the Vaisala USB cable and a

terminal application, see section Connecting With a on page 34.

For a description of the serial commands, see section Calibration and

Adjustment Commands on page 53.

1.

Connect the GMW90 series transmitter to your computer using the

Vaisala USB cable (order code 219690).

2.

Start a terminal application and open a new session to the service

port of the transmitter. The serial line settings are 19200, N, 8, 1.

3.

Before changing the adjustment, issue the following commands to

see the transmitter’s current adjustment information:

pass 9000

cco2

ct

crh

ctext

cdate

Note that the CRH command is not available if your transmitter

does not support humidity measurement.

4.

Place the entire transmitter in the desired reference condition and

allow the measurement to stabilize. Follow the stabilization from

the serial line (output from the R command) or the display.

5.

You can now use the CCO2, CT, and CRH commands to adjust

the transmitter’s measurement. Refer to the command descriptions

for the available options.

6.

After performing the adjustment, verify from the serial line or the

display that the measurement has been corrected.

7.

After completing the adjustments, you can enter a descriptive text

string in the transmitter’s memory using the CTEXT command,

and note the adjustment date using the CDATE command.