Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 55

Clear User Adjustment of CO

2

Measurement

CCO2 [RESET]<cr>

Example:

>pass 9000

>cco2 reset

OK

Adjust Humidity Measurement

Use the CRH command to perform a humidity adjustment of the relative

humidity (RH) measurement. You can do a 1-point or a 2-point

adjustment, or clear the adjustment information from the HTM10

module. Note that the factory adjustment will remain intact when user

adjustment is cleared.

NOTE

Before using the CRH command, read section Notes for RH Adjustment

on page 65.

Show Current RH Adjustment

CRH<cr>

Example (showing default offset and gain):

>pass 9000

>crh

RH Gain : 1.000

RH Offset : 0.000