background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 51

Table 9

FORM Command Parameters

Measured Parameter

Abbreviation in FORM Command

Carbon dioxide

CO2

Relative humidity

RH

Temperature

T

Dew/frostpoint temperature

Tdf

Dewpoint temperature

Td

Wetbulb temperature

Tw

Enthalpy

h

Mixing ratio

x

Absolute humidity

a

Dew/frostpoint depression

dTd

Table 10

FORM Command Modifiers

Modifier

Description

x.y

Length modifier (number of digits and decimal places)

#t

Tabulator

#r

Carriage-return

#n

Line feed

""

String constant, length 1 ... 15 characters

#xxx

ASCII code value (decimal) of a special character;

for example,

#027 for ESC

Ux

Shows the name of the measurement unit using “x” number

of characters. For example,

U3 shows the name of the

measurement unit with three characters

CS2

Modulus-256 checksum of message sent so far, ASCII

encoded hexadecimal notation

CS4

Modulus-65536 checksum of message sent so far, ASCII

encoded hexadecimal notation

CSX

NMEA xor-checksum of message sent so far, ASCII encoded

hexadecimal notation

NOTE

When entering modifiers, you can also use the backslash character “\”

instead of the hash “#”.