User's Guide _______________________________________________________________________

58 ___________________________________________________________________ M211659EN-B

Clear User Adjustment of T Measurement

CT [RESET]<cr>

Example:

>pass 9000

>ct reset

OK

Enter Calibration and Adjustment

Information

Use the CTEXT command to store a text string that describes the

calibration and/or adjustment. To enter a text string with spaces, enclose

the string in quotation marks. Use the CDATE to store the date.

CTEXT [text]<cr>

CDATE [YYYY-MM-DD]<cr>

Examples:

>pass 9000

>ctext “adjusted rhlab/Tech021”

“adjusted rhlab/Tech021”

>cdate 2011-12-08

Calibration date : 2011-12-08