Chapter 3 ________________________________________________________________ Installation

VAISALA ________________________________________________________________________ 27

Wiring GMW93R/RA

Recommended wiring for long cables:

1402-032

Figure 13

Wiring GMW93R/RA

-Vs terminal is internally connected to GND terminal, so you can also

use the -Vs terminal as common ground. Maximum cable resistance is

2.5

Ω (24V supply, 0 ... 10 V output):

1402-031

Figure 14

Alternate Wiring for GMW93R/RA

Wiring GMW93

GMW93 is wired in the same way as GMW93R/RA, except for the

humidity output that is not present:

1402-030

Figure 15

Wiring GMW93

T+

GND

-Vs

CO

2

+

+Vs

Power supply

18 ... 35 VDC

or 24 VAC ±20%

V

R

L

= 10 kΩ min.

V

RH+

V

T+

GND

-Vs

CO

2

+

+Vs

Power supply

18 ... 35 VDC

or 24 VAC ±20%

R

L

= 10 kΩ min.

V

RH+

V

V

T+

GND

-Vs

CO

2

+

+Vs

Power supply

18 ... 35 VDC

or 24 VAC ±20%

V

R

L

= 10 kΩ min.

V