Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 63

CHAPTER 5

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

the GMW90 series.

Cleaning

The body of the transmitter can be cleaned by wiping with a moistened

lint-free cloth. Do not use cleaning agents or solvents, or blow

pressurized air into the transmitter housing.

Do not attempt to clean contaminated measurement modules. Dirty

modules should always be replaced with new calibrated modules.

Calibration and Adjustment

GMW90 series transmitters are fully calibrated as shipped from factory.

GMW90 series transmitters have a display that makes it easy to compare

the measured readings against any portable calibration reference. Note

that depending on the ordered configuration, the display may be hidden

under the sliding cover.

Adjustment Methods

You can adjust the measurements in the following ways:

- 1-point adjustment using the trimmers under the sliding cover. See

section Adjustment Using Display and Trimmers on page 66.

- 1-point or 2-point adjustment using the service port. See the following

sections:

- Adjustment Using a Hand-Held Meter on page 67.
- Adjustment Using a Computer on page 68.

If adjustment is not enough to restore the measurement accuracy of the

transmitter, you can also replace the measurement modules. See section

Repair Maintenance on page 69.