Error messages, O table 12 on, Table 12 error messages – Vaisala GMW90 User Manual

Page 75

Advertising
background image

Chapter 6 ____________________________________________________________ Troubleshooting

VAISALA ________________________________________________________________________ 73

Error Messages

Table 12

Error Messages

Error Text

on Display

Error ID on

Serial Line

Possible Cause Remedy

HTM10 01

23

Communication

problem with

HTM10 module.

Check that the HTM10

module sits firmly in place.

Remove and reconnect.

HTM10 04

21

Problem with

humidity

measurement.

- Check for missing or

damaged HUMICAP

®

sensor on the HTM10

module.

- Check for condensation

on the HUMICAP

®

sensor. Wait for the

sensor to dry out.

HTM10 03

4

Internal problem

with HTM10

module.

- Restart the transmitter.

- Replace the module if

unable to remove the

problem.

HTM10 05

22

HTM10 06 42
GM10 01

79

Communication

problem with

GM10 module.

Check that the GM10 module

sits firmly in place. Remove

and reconnect.

GM10 03

81

Internal problem

with GM10

module.

- Restart the transmitter.

- Replace the module if

unable to remove the

problem.

GM10 04

82

GM10 10

28

GM10 11

88

GM10 12

89

Dirty or damaged

GM10 module.

Replace module.

Internal 1

1

Internal problem

with the

transmitter.

- Restart the transmitter.

- Restore the factory

settings using service

port or DIP switches if

reset does not help.

Internal 2

2

Internal 3

3

Internal 4

43

Advertising