Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 100

Advertising
background image

Dakota Ultrasonics

96

11.7 Zero

The DFX-8 has a feature to manually adjust the electronic zero. This feature used in
conjunction with manually adjusting the velocity, are the primary features necessary
to perform a ‘crude’ manual screen calibration. The procedure for adjusting the zero
is listed below:

Adjusting Zero

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until ZERO is highlighted.

3) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the value. When the correct zero value is being

displayed, proceed to step 8.

4) Alternatively, press the ENTER key to display the Digits Edit Box.

5) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.