Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 50

Advertising
background image

Dakota Ultrasonics

46

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

move the gate over the waveform to be measured, as demonstrated in the

screen shot above.

Note: In order to trigger on a signal, the waveform must be inside the

boundaries of the gate.

Calibrating the First Point

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right and the ESC key multiple times to tab left until

the CAL menu is highlighted and displaying the submenu items, as shown in

the screenshot above. Alternatively, press the

LEFT and RIGHT arrow

keys multiple times using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until MATL 1PT is highlighted.

3) Press the ENTER key to display the Digits Edit Box.

4) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

5) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

6) Repeat steps 4 & 5 until the known thickness value is correctly displayed.