Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 59

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

55

When TRIG mode is enabled, the calculated measurements will be displayed below
the waveform display area at (A), and the signal amplitude as a percentage of full
screen height at (B).

Curved Surface Correction:

The DFX-8 will also compensate for convex curved surfaces of longitudinally welded
pipes. A curved surface will produce a greater sound path than a flat surface will,
using a transducer with a given angle. The user should consider the angle used on
curved surfaces, with respect to the material thickness, in order to be sure a
reflection will occur from the opposite surface of the test material. In order to utilize
the curved surface compensation, the outside diameter must be entered into the
DFX-8.

6.3 Setup Trigonometry Mode

This section provides a step by step procedure to enable and setup the TRIG mode.
Note: A two point calibration should be done prior to proceeding. Refer to Chapter
Five for the calibration procedures.

For quick access and editing, refer to navigating the Hot Menus page 22.

Advertising
This manual is related to the following products: