Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 79

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

75

1) With all the parameters set in the previous section, press the MEAS key

once to return to the main measurement screen.

Note: The user should now see a highlighted cursor displayed in the waveform

area, starting at the left most side of the display area.

ex. 1

ex. 2

2) Press the left button wheel, press the

LEFT and RIGHT arrow keys to

scroll through the 9 individual Hot Menus, and the

UP and DOWN arrow

keys to scroll through the Hot Menu items until the TCG BOX: cell is

highlighted. Now that the TCG BOX: cell is highlighted, the cursor can be

controlled by moving the

LEFT and RIGHT

arrow keys on the right

button wheel.

3) Couple the transducer to the first calibration standard, and peak up on the

reflector to achieve maximum signal amplitude.

4) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

move the cursor over the waveform to be used as the first point in the TCG

curve, and press the ENTER key to set the first point. Illustrated in ex.1

above.

5) Repeat steps 3 & 4 to set multiple points as is necessary. Illustrated in ex.2

above.