Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 45

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

41

2) Press the ENTER key to display the Digits Edit Box.

3) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

4) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

5) Repeat steps 3 & 4 until the desired range value is displayed.

6) Press the OK key to set the value and return to the measure screen.

Note: Our range calibration range for this example is 4 inches. Therefore

setting a range great than 4 inches will enable the user to view the entire range

on the display.

Enabling and Setting Up a Gate

Refer to the illustrations above. In order to measure, a gate is required. The

width of the gate should be small enough, in terms of its width, in order to select

specific reflections to measure. The screenshot above shows a gate enabled to

select the 1 inch calibration standard, and in the second the 4 inch calibration

standard. For quick access and editing, refer to navigating the Hot Menus page

22.