Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 51

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

47

7) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel the one point calibration.

8) Finally, press the MEAS key to return to the Measurement screen.

Apply a drop of couplant on the transducer and place the transducer in steady

contact with the 4” calibration standard. Be sure that the reading is stable and

the repeatability indicator, in the top left corner of the display, is fully lit and

stable. Also, be sure there is sufficient signal amplitude in order to avoid ‘peak

jumping’ or missing the first cycle.

Adjusting the Gate

1) Press the MEAS key once to activate the measure menu items. Press the

MEAS key multiple times to move right and the ESC key multiple times to

move left, until the GATE1 cell is highlighted. Alternatively, press the

LEFT and RIGHT arrow keys multiple times using the left button wheel.

2) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

move the gate over the waveform to be measured, as demonstrated in the

screen shot above.

Note: In order to trigger on a signal, the waveform must be inside the

boundaries of the gate.

Calibrating the Second Point