until the AWS menu is highlighted and displaying the submenu items.

Alternatively, press the

LEFT and RIGHT arrow keys multiple times

using the left button wheel.

2) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until CURVE is highlighted.

3) Press the

LEFT and RIGHT

arrow keys on the right button wheel to

toggle the CURVE feature ON/OFF. The menu items should now become

active and adjustable, as illustrated above.

Enter % Screen Height

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until % SCRN HEIGHT is highlighted, followed by pressing the

LEFT and RIGHT

arrow keys on the right button wheel to scroll the value.

When the correct height is displayed, proceed to the next setting.

2) Alternatively, press the ENTER key to display the Digits Edit Box.

3) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

4) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

5) Repeat steps 3 & 4 until the height is correctly displayed.

6) Press the OK key to set the % SCRN HEIGHT and return to the menu

screen, or ESC to cancel entering the % SCRN HEIGHT.

Note: % Screen Height is an arbitrary level, and is not specified by the AWS

D1.1. The user should set a value that will optimize viewing defects that are

larger and smaller than the reference level. Typically this level is set between

50%-80% full screen height.