Dakota Ultrasonics DFX-8 plus MANUAL1 User Manual

Page 71

Advertising
background image

DFX-8 Series Ultrasonic Flaw Detectors

67

%FSH – The top of the waveform display represents 100% full screen
height. Any waveform amplitudes generated greater than 100% full
screen height will not be displayed. 100% is the greatest value that can
be displayed.

Note: The MEAS option selected is overridden with %FSH while the
DAC curve is in DRAW mode. Once the DRAW mode is turned off, the
select MEAS option will be displayed.

 The

MEAS, CURVE, and TRIGGER options can be changed at any time once

the DAC curve has been created.

 In order to move the highlighted cursor over a waveform to use as a point in

generating the DAC curve, the “DAC BOX:” hot menu item in the main
measurement screen must be highlighted.

8.3 Creating a DAC Curve

This section provides a step by step procedure to create a DAC curve. Note: a two
point calibration should be done prior to proceeding. Refer to Chapter Five for the
calibration procedures. Also, if the TRIG function will be used, it should be setup
prior to proceeding. Refer to Chapter Six for the setup procedure.

For quick access and editing, refer to navigating the Hot Menus page 22.

Activating & Setting the Parameters

Activating DAC

1) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the DAC menu is highlighted and displaying the submenu items.

Advertising
This manual is related to the following products: