DFX-8 Series Ultrasonic Flaw Detectors

37

1) Press the

UP and DOWN arrow keys multiple times on the left button

wheel until THRESHOLD1 is highlighted, followed by pressing the

LEFT and RIGHT arrow keys multiple times using the left button wheel to

scroll the value. When the correct threshold level is displayed, proceed to

the next menu item setting.

2) Alternatively, press the ENTER key to display the Digits Edit Box.

3) Press the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll the highlighted value.

4) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to scroll the digit locations.

5) Repeat steps 3 & 4 until the threshold level is correctly displayed.

6) Press the OK key to set the THRESHOLD1 level and return to the menu

screen, or ESC to cancel entering the THRESHOLD1 level.

Note: This sets the overall sensitivity level of the gate which is the threshold.

Select Gate Polarity

1) Use the

UP and DOWN arrow keys multiple times on the left button

wheel to scroll through the sub menu items until POLARITY1 is highlighted.

2) Press the

LEFT and RIGHT arrow keys multiple times using the left

button wheel to toggle the POLARITY feature positive or negative.

3) Press the MEAS key to return to the measurement screen.

Note: The GATE1 and WIDTH1 values can now be adjusted or tuned directly

from the Hot Menus on the main measurement screen.

Note: Sets where the detection will occur, on the negative or positive going

echoes.